製品名
Nitrogen (total) Cell Test, photometric, 0.5-15.0 mg/L (N), Spectroquant®
Quality Level
product line
Spectroquant®
usage
sufficient for 25 tests
specific analyte(s)
nitrogen
measuring range
0.5-15.0 mg/L (N)
technique(s)
photometry: suitable
application(s)
agriculture
aquaculture
boiler and cooling water
disposal drainage water
environmental
ground and surface water
milk and dairy products
wastewater
compatibility
for use with Spectroquant® Nova 60 A, for use with Spectroquant® Prove 100 (plus), for use with Spectroquant® Prove 300 (plus), for use with Spectroquant® Prove 600 (plus)
detection method
photometric (Koroleff digestion, 2,6-dimethylphenol)
storage temp.
15-25°C
General description
The method corresponds to ISO 23697-1, a norm standardizing the use of this "small-scale sealed tube method" for this important parameter. The digestion is analogous to EN ISO 11905-1. The determination of nitrate is analogous to DIN 38405-9.
This Spectroquant® Nitrogen (total) Cell Test allows the reliable determination of total nitrogen in water samples. Total nitrogen is the sum of total Kjeldahl nitrogen (organic nitrogen and ammonia), nitrate, and nitrite. Prior to the determination the sample needs to be digested in order to transfer organic and inorganic nitrogen containing compounds into nitrate that further reacts to a photometrically analyzable compound.
The Spectroquant® Cell Tests come with prefilled 16 mm round cells and all the required reagents to perform the analysis according to the instruction leaflet provided.
All Spectroquant® Cell and Reagent Test Kits are equipped with the unique Live ID (2D barcode) which allows seamless method recognition and contains essential information such as lot number, expiry date, and automatic calibration updates.
Legal Information
1 of 1
当該品目 | |||
|---|---|---|---|
| detection method photometric (Koroleff digestion, 2,6-dimethylphenol) | detection method photometric (Koroleff digestion, 2,6-dimethylphenol) | detection method photometric (Koroleff digestion, nitrospectral) | detection method photometric (2,6-Dimethylphenol) |
| specific analyte(s) nitrogen | specific analyte(s) nitrogen | specific analyte(s) nitrogen | specific analyte(s) nitrate |
| technique(s) photometry: suitable | technique(s) photometry: suitable | technique(s) photometry: suitable | technique(s) photometry: suitable |
| measuring range 0.5-15.0 mg/L (N) | measuring range 10-150 mg/L (N) | measuring range 0.5-15.0 mg/L (N) | measuring range 102-996 mg/L (NO3-), 23-255 mg/L (NO3-N) |
| compatibility for use with Spectroquant® Nova 60 A, for use with Spectroquant® Prove 300 (plus), for use with Spectroquant® Prove 100 (plus), for use with Spectroquant® Prove 600 (plus) | compatibility for use with Spectroquant® Nova 60 A, for use with Spectroquant® Prove 100 (plus), for use with Spectroquant® Prove 300 (plus), for use with Spectroquant® Prove 600 (plus) | compatibility for use with Spectroquant® Move 100, for use with Spectroquant® Nova 60 A, for use with Spectroquant® Prove 100 (plus), for use with Spectroquant® Prove 300 (plus), for use with Spectroquant® Prove 600 (plus) | compatibility for use with Spectroquant® Nova 60 A, for use with Spectroquant® Prove 100 (plus), for use with Spectroquant® Prove 300 (plus), for use with Spectroquant® Prove 600 (plus) |
| usage sufficient for 25 tests | usage sufficient for 25 tests | usage sufficient for 25 tests | usage sufficient for 25 tests |
signalword
Danger
Hazard Classifications
Acute Tox. 4 Oral - Eye Dam. 1 - Flam. Liq. 3 - Met. Corr. 1 - Ox. Liq. 3 - Resp. Sens. 1 - Skin Corr. 1A - Skin Sens. 1 - STOT SE 3
target_organs
Respiratory system
保管分類
5.1B - Oxidizing hazardous materials
flash_point_f
84.2 °F
flash_point_c
29 °C
適用法令
試験研究用途を考慮した関連法令を主に挙げております。化学物質以外については、一部の情報のみ提供しています。 製品を安全かつ合法的に使用することは、使用者の義務です。最新情報により修正される場合があります。WEBの反映には時間を要することがあるため、適宜SDSをご参照ください。
Please refer to KIT Component information
pdsc
Please refer to KIT Component information
prtr
Please refer to KIT Component information
fsl
Please refer to KIT Component information
ishl_indicated
Please refer to KIT Component information
ishl_notified
Please refer to KIT Component information
cart
キットコンポーネントの情報を参照してください
jan
試験成績書(COA)
製品のロット番号・バッチ番号を入力して、試験成績書(COA) を検索できます。ロット番号・バッチ番号は、製品ラベルに「Lot」または「Batch」に続いて記載されています。
プロトコル
Soil analysis to determine nitrogen (total) content using Spectroquant® Nitrogen Test Kit and Spectroquant® Prove spectrophotometers by phosphorus molybdenum blue method.
資料
全窒素分析キットは、水や廃水のテストのための迅速かつ正確で費用対効果の高い方法です。
Water quality can impact the accuracy of the Kjeldahl method. Avoid contaminants and assure results by selecting the optimal water purification solution.
Explore the Total Nitrogen Analysis kit for rapid, precise measurement of total nitrogen in water and wastewater, supporting reliable environmental monitoring.
グローバルトレードアイテム番号
| カタログ番号 | GTIN |
|---|---|
| 1006130001 | 04022536733821 |