HPLC Analysis of Proteins on BIOshell A400 Protein C18
Materiais
coluna analítica
Nº do produto
Descrição
Preços
padrão
Nº do produto
Descrição
Preços
Trypsin Inhibitor from Glycine max (soybean)
BioUltra, lyophilized powder, ≥95% (Kunitz inhibitor, SDS-PAGE)CONDITIONS
column
BIOShell A400 C18, 10 cm x 2.1 mm I.D., 3.4 μm particles (67463-U)
mobile phase
[A] (water, 0.1% TFA); (acetonitrile, 0.08% TFA), 75:25; [B] (water, 0.1% TFA); (acetonitrile, 0.08% TFA), 25:75
gradient
20 to 100% B in 8 min
flow rate
0.3 mL/min
pressure
2480 psi total pressure (193 bar)
column temp.
30 °C
detector
UV, 215 nm
injection
2 μL
sample
lysozyme, 0.06; haptoglobulin, 0.06; conalbumin, 0.12; tryspin inhibitor, 0.18; glutamate dehydrogenase, (0.18; ß-galactosidase, 0.12; serum albumin chicken, 0.12 g/L in mobile phase A)
Descrição
Nota de análise
Analysis of proteins of various hydrophobicity on BIOshell A400 Protein C18 to compare the selectivity and retention times.
Informações legais
null