GC Analysis of Volatiles in White Wine on VOCOL® after SPME using a 75 μm Carboxen/PDMS Fiber

GC Analysis of Volatiles in White Wine on VOCOL® after SPME using a 75 μm Carboxen/PDMS Fiber suitable for GC, application for SPME

Materiais

Fibra para SPME

Nº do produto
Descrição
Preços

Conjunto de fibras de SPME Carboxeno/polidimetilsiloxano (CAR/PDMS)

df 75 μm(CAR/PDMS, for use with manual holder, needle size 24 ga

coluna analítica

Nº do produto
Descrição
Preços

VOCOL® Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 1.50 μm

componente de fase móvel

Nº do produto
Descrição
Preços

Methyl formate

analytical standard

padrão

Nº do produto
Descrição
Preços

Hexyl acetate

analytical standard

Octanoic acid

analytical standard

Ethyl butyrate

analytical standard

Isoamyl alcohol

≥98%, FG

2-Methyl-1-butanol

≥99%

Sulfur dioxide

≥99.9%

Ethyl hexanoate

analytical standard

Ethyl octanoate

analytical standard

2,3-Butanediol

analytical standard, mixture of racemic and meso forms

produto relacionado

Nº do produto
Descrição
Preços

Suporte de fibra para SPME

for use with manual sampling

Suporte de fibra para SPME

for use with CTC CombiPAL and Gerstel MPS2

CONDITIONS

sample/matrix

white wine + 25% NaCl

SPME fiber

Carboxen/PDMS, 75 μm (57318)

extraction

10 min, 40 °C (headspace)

desorption process

3 min at 290 °C

column

VOCOL, 30 m × 0.25 mm I.D., 1.5 μm (24205-U)

oven

40 °C (2 min) to 220 °C at 10 °C/min

inj. temp.

290 °C

detector

GC-MS, quadrupole, m/z = 31-240

carrier gas

helium, 35 cm/sec

injection

splitless/split, closed 1 min

liner

0.75 mm liner

sample

white wine

Descrição

Nota de análise

In this application, SPME was used for the analysis of volatiles in white wine using a headspace method. The presence of aroma compounds in white wine is one of the most important attributes of its final quality. A Carboxen/PDMS fiber was inserted into the headspace of the vial for 10 minutes at 40 °C and analyzed by GC/MS. The thick film of the VOCOL GC column helps focus the volatile analytes.