Przejdź do zawartości
Merck

HPLC Analysis of Protein Standards on Zenix® SEC-300 at Low Flow Rate

HPLC Analysis of Protein Standards on Zenix® SEC-300 at Low Flow Rate application for HPLC

Materiały

kolumna analityczna

Numer produktu
Opis
Cennik

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 7.8 mm, 3 μm particle size, stainless steel

składnik fazy ruchomej

Numer produktu
Opis
Cennik

Phosphoric acid

≥85 wt. % in H2O, ≥99.999% trace metals basis

standardowe

Numer produktu
Opis
Cennik

Myoglobin from equine heart

≥90% (SDS-PAGE), essentially salt-free, lyophilized powder

Uracil

≥99.0%

Thyroglobulin from bovine thyroid

powder, ≥90% (agarose gel electrophoresis)

IgG from rat serum

reagent grade, ≥95% (SDS-PAGE), essentially salt-free, lyophilized powder

Bovine Serum Albumin

lyophilized powder, crystallized, ≥98.0% (GE)

CONDITIONS

column

Zenix SEC-300, 30 cm x 7.8 mm I.D., 3 μm (Z777033)

mobile phase

0.15 M potassium phosphate monobasic, pH 7.0 with phosphoric acid

flow rate

0.1 mL/min

pressure

173 psi (12 bar)

column temp.

25 °C

detector

UV, 215 nm

injection

4 μL

sample

protein standards

Opis

Komentarz do analizy

Due to the low diffusion coefficents of proteins, optimal flow in size exclusion chromatography is often very low. However, the disadvantage is long run times. This example demonstrates both features.

Informacje prawne

Zenix is a registered trademark of Sepax Technologies