HPLC Analysis of Sugars by HILIC Chromatography using Ascentis® Express OH-5 Column with ELSD Detection

HPLC Analysis of Sugars by HILIC Chromatography using Ascentis® Express OH-5 Column with ELSD Detection application for HPLC

Materiály

Analytical column

Č. produktu
Popis
Stanovení ceny

Ascentis® Express OH5, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 15 cm × 4.6 mm

CONDITIONS

column

Ascentis Express OH-5, 15 cm x 4.6 mm, I.D., 2.7 μm particles (53778-U)

mobile phase

[A] 0.005 M ammonium formate, pH 3.15 with concentrated TFA [B] acetonitrile (30:70, A:B)

flow rate

0.5 mL/min

pressure

841 psi (58 bar)

column temp.

35 °C

detector

ELSD (detector other: Nebulizer 60º C, Evaporator 80º C, Gas flow 1.60 SLM (Standard Liters/ Minute), LED 100%, Smoothing 5.0 S, PMT ( Photomultiplier) Gain 1.0)

injection

10 μL

sample

0.5 mg/mL each prepared in 50:50 Water:Acetonitrile

Popis

Analysis Note

This application shows the analysis of sugars by HILIC Chromatography with ELSD detection. The purpose of the work shown here was to develop a HILIC method that could quickly separate mono- and di-saccharides with good sensitivity and reproducibility utilizing the Ascentis Express OH-5 column and an Evaporative Light Scattering Detecctor (ELSD) detector.