Analytical column

Č. produktu
Popis
Stanovení ceny

Equity-1 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 1.00 μm

related product

Č. produktu
Popis
Stanovení ceny

Carbotrap® 300

glass TD tube, fritted, O.D. × L 1/4 in. × 3 1/2 in., unconditioned, pkg of 10 ea

CONDITIONS

thermal desorption tube

Carbotrap® 300 glass TD tube, fritted, O.D. × L 1/4 in. × 3 1/2 in., unconditioned, pkg of 10 ea (30141-U)

sample/matrix

TO15/17 Calibration Mix, 1 PPM in N2 (41973-U)

dry purge

1 min

bake

350 °C

transfer line/valve temp.

175 °C

column

FS CAP Equity-1 30 m x 0.25 mm I.D. x 1.0 μm (28049-U)

oven

40 °C/8 min to 100 °C @ 5 °C/min to 220 °C @ 15 °C/min, hold for 10 min

detector

MSD

MSD interface

175 °C

scan range

35-265

carrier gas

Helium