Přejít k obsahu
Merck

OSHA Method 52: GC Analysis of Formaldehyde on SUPELCOWAX 10 after Collection/Desorption using ORBO-24

OSHA Method 52: GC Analysis of Formaldehyde on SUPELCOWAX 10 after Collection/Desorption using ORBO™-24 suitable for GC, application for air monitoring

Materiály

analytical column

Č. produktu
Popis
Stanovení ceny

SUPELCOWAX 10 Capillary GC Column

L × I.D. 30 m × 0.53 mm, df 0.50 μm

related product

Č. produktu
Popis
Stanovení ceny

ORBO 24 2-HMP on Amberlite XAD®-2 (20/40), 150/75 mg

W,W,W separators, O.D. × L 6 mm × 105 mm, pkg of 25 ea

standard

Č. produktu
Popis
Stanovení ceny

Formaldehyde solution

stabilized with methanol, ~37 wt. % in H2O, certified reference material

CONDITIONS

sample/matrix

HMP-coated Supelpak 20 spiked with formaldehyde (10 μg)

adsorbent tube

ORBO-24 (20231)

column

SUPELCOWAX 10, 30m × 0.53 mm I.D., 0.5 μm (25325)

oven

50 °C (2 min) to 200 °C at 10 °C/min

detector

NPD, 220 °C

carrier gas

helium, 7 mL/min

injection

1 μL direct, 220 °C

Popis

Legal Information

ORBO is a trademark of Sigma-Aldrich Co. LLC