Pular para o conteúdo
Merck

HPLC Analysis of Amphetamine and Methamphetamine Enantiomers on Astec® CHIROBIOTIC® V2

HPLC Analysis of Amphetamine and Methamphetamine Enantiomers on Astec® CHIROBIOTIC® V2 application for HPLC

Materiais

coluna analítica

Nº do produto
Descrição
Preços

Astec® CHIROBIOTIC® V2 Chiral HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

padrão

Nº do produto
Descrição
Preços

(+)-Methamphetamine hydrochloride

D-Amphetamine hemisulfate salt solution

1.0 mg/mL in methanol, analytical standard, for drug analysis

(+)-Methamphetamine hydrochloride solution

analytical standard, for drug analysis, 1.0 mg/mL in methanol

CONDITIONS

column

CHIROBIOTIC V2, 25 cm x 4.6 mm I.D., 5 μm particles (15024AST)

mobile phase

[A] methanol; [B] acetic acid; [C] ammonium hydroxide; (99.88:0.1:0.01, A:B:C)

flow rate

1 mL/min

column temp.

25 °C

detector

UV, 210 nm

injection

1 μL

sample

1 mg/mL in methanol

Descrição

Informações legais

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC