GC Analysis of Volatiles in Milk (Before and After Exposure to Sunlight) on Supel-Q PLOT after SPME using a 75 μm PDMS/Carboxen Fiber

GC Analysis of Volatiles in Milk (Before and After Exposure to Sunlight) on Supel-Q PLOT after SPME using a 75 μm PDMS/Carboxen Fiber suitable for GC, application for SPME

Materiais

Fibra para SPME

Nº do produto
Descrição
Preços

Conjunto de fibras de SPME Carboxeno/polidimetilsiloxano (CAR/PDMS)

df 75 μm(CAR/PDMS, for use with manual holder, needle size 24 ga

coluna analítica

Nº do produto
Descrição
Preços

Supel-Q PLOT Capillary GC Column

L × I.D. 30 m × 0.32 mm, average thickness 15 μm

padrão

Nº do produto
Descrição
Preços

4-metil-2-pentanona

analytical standard

Hexanal

analytical standard

Dimethyl disulfide

analytical standard

produto relacionado

Nº do produto
Descrição
Preços

Suporte de fibra para SPME

for use with manual sampling

Suporte de fibra para SPME

for use with CTC CombiPAL and Gerstel MPS2

CONDITIONS

sample/matrix

3 g of 2% milk + 10 μL internal standard solution (20 μg/mL 4-methyl-2-pentanone) (9 mL GC vial)

SPME fiber

PDMS/Carboxen, 75 μm (57318)

extraction

headspace, 15 min with constant stirring at 45 °C

desorption process

5 min, 250 °C

column

Supel-Q PLOT, 30 m × 0.32 mm I.D. (24242)

oven

70 °C (2 min) to 140 °C at 6 °C/min (2 min hold) then to 220 °C at 6 °C/min (5 min hold)

inj. temp.

250 °C

detector

GC-MS ion trap, m/z = 33-300

carrier gas

helium, 35 cm/sec

injection

splitless, closed 2 min

liner

0.75 mm liner

sample

Milk

Descrição

Nota de análise

In this study, SPME was used for the analysis of volatiles in milk using a headspace method. Light can catalyze the formation of peroxides that can cleave double bonds in fatty acids. The cleaved by-products form aldehydes that can give the food a pungent off flavor. Dr. Ray Marsili investigated the feasibility of using SPME as a substitute sample preparation device prior to GC/MS analysis. The Carboxen/PDMS fiber was evaluated for the extraction of low ppb levels of pentanal, hexanal, and dimethyldisulfide. The results of the comparison of SPME to dynamic headspace analysis indicate that SPME-GC/MS is more accurate and precise while providing sensitivities equivalent to DH-GC/MS. The GC column used, Supel-Q PLOT, operates via gas-solid chromatography with a DVB adsorbent and permits the separation of a wide range of analytes.