Przejdź do zawartości
Merck
Wszystkie zdjęcia(2)

Dokumenty

1.09174

Millipore

MAS-100 Iso NT® Remote Microbial Air Sampler

with USB, digital I/O and Ethernet interfaces

Synonim(y):

MAS-100 Iso NT® Control unit

Zaloguj sięWyświetlanie cen organizacyjnych i kontraktowych


About This Item

Kod UNSPSC:
41103403
NACRES:
NA.74

Materiały

aluminum hardware

agency

according to ISO 14698

Właściwości

installed

opakowanie

pkg of 1 kit

producent / nazwa handlowa

MAS-100 Iso NT®

Zastosowanie

air monitoring
pharmaceutical

kompatybilność

for use with 90 mm Petri dishes

temp. przechowywania

no temp limit

Opis ogólny

The MAS-100 ISO NT® is designed to monitor the microbial contamination of the air in isolators. Like all other MAS-100 air samplers, it′s based on the Andersen impaction principle. The control unit with electronic and moving blower parts can be located outside the critical zone, while the single sampling head for standard 90 to 100 mm Petri dishes collect samples in the containment zone. The tubing length can be up to 10 m when used with VHP decontamination.
The innovative double valve system enables the sampling heads to be integrated into the decontamination process with VHP. Each unit has both an air and a sterilization pump that operate autonomously. The instrument can be operated via an LCD screen.
The fully automatic calibration of the device guarantees maximum efficiency and accuracy. The system measures the air flow and regulates it to a constant value of 100 liters/min. These systems have the highest airflow accuracy of ±2.5%, compared to others that can be as high as ±10%. The instruments are designed and built according to GAMP 4 and are independently validated according to ISO 14698-1 with very good physical efficiency providing a D50 value below 1.1μm and a high biological efficiency result. The efficiencies and the D50 values are independent of the tube length and are comparable to the other members of the MAS-family.
The MAS-100 ISO NT® is equipped with one valve for one sampling head and RS232, USB and digital IN/OUT interfaces for communication. In case another communication interface is required MAS-100 ISO NT® Control unit with Profibus with Profibus interface is available. The type of sample base (standard or easy clean version) as well as the perforated lids can be selected separately according to the specific needs of the sample location and workflow.

Zastosowanie

MAS-100 ISO NT® Control unit with Ethernet is suitable for microbial monitoring in isolators and cleanrooms. The microbial air sampling devices are extensively used in Pharmaceutical applications.

Komponenty

Control unit equipped with display and valve connection (including cable) for one sampling head, 2 Tri-Clamps, 2 Silicon gaskets, a range of phoenix connectors, Ethernet module, USB cable, digital I/O, USB stick with all manuals and documentation.

Informacje prawne

MAS-100 ISO NT is a registered trademark of MBV AG

produkt powiązany

Numer produktu
Opis
Cennik

wyposażenie dodatkowe

Numer produktu
Opis
Cennik

Kod klasy składowania

10-13 - German Storage Class 10 to 13


Certyfikaty analizy (CoA)

Poszukaj Certyfikaty analizy (CoA), wpisując numer partii/serii produktów. Numery serii i partii można znaleźć na etykiecie produktu po słowach „seria” lub „partia”.

Masz już ten produkt?

Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.

Odwiedź Bibliotekę dokumentów

Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.

Skontaktuj się z zespołem ds. pomocy technicznej