Pular para o conteúdo
Merck

HPLC Analysis of Protein Standards on Zenix® SEC-300 by Size Exclusion As Affected by Potassium Chloride

HPLC Analysis of Protein Standards on Zenix® SEC-300 by Size Exclusion As Affected by Potassium Chloride application for HPLC

Materiais

Padrão

Nº do produto
Descrição
Preços

Myoglobin from equine heart

≥90% (SDS-PAGE), essentially salt-free, lyophilized powder

Uracil

≥99.0%

Thyroglobulin from bovine thyroid

(For Gel Filtration Chromatography)

IgG from rabbit serum

reagent grade, ≥95% (SDS-PAGE), essentially salt-free, lyophilized powder

Albumina sérica bovina

lyophilized powder, ≥96% (agarose gel electrophoresis)

acessório

Nº do produto
Descrição
Preços

EXP® Pre-Column Filter

holder with EXP titanium hybrid ferrule, pack of 1 sets

EXP® Pre-Column Filter Cartridges

0.5μm, pkg of 5 ea

coluna analítica

Nº do produto
Descrição
Preços

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 4.6 mm, 3 μm particle size, stainless steel

componente de fase móvel

Nº do produto
Descrição
Preços

Cloreto de potássio

BioXtra, ≥99.0%

CONDITIONS

column

Zenix SEC-300, 30 cm x 4.6 mm I.D., 3 μm particles (Z777028)

mobile phase

[A], 0.15 M potassium phosphate, monobasic, pH 7.0 with potassium hydroxide; [B] 0.15 M potassium phosphate, monobasic, 0.2 M potassium chloride, pH 7.0 with potassium hydroxide; (mobile phase is either 100% A or 100% B)

flow rate

0.3 mL/min

pressure

1000 psi (69 bar)

column temp.

25 °C

detector

UV, 215 nm

injection

2.5 μL

sample

Waters BEH200 SEC protein standard

Descrição

Nota de análise

In size exclusion chromatography, addition of halide salts (typically sodium or potassium chloride) in modest concentration, to an aqueous buffer, can yield improved chromatographic performance with protein analytes. In this case, as regards chromatographic performanc, little impact is noted except in the case of IgG, where peak shape is improved.