Pular para o conteúdo
Merck

GC Analysis of FAMEs in Atlantic Salmon on Omegawax

GC Analysis of FAMEs in Atlantic Salmon on Omegawax suitable for GC

Materiais

coluna analítica

Nº do produto
Descrição
Preços

Omegawax Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

CONDITIONS

column

Omegawax, 30 m x 0.25 mm I.D., 0.25μm (24136)

oven

170 °C, 1 °C/min to 225°C

inj. temp.

250 °C

detector

FID, 260 °C

carrier gas

helium, 1.2 mL/min

injection

1 μL, 100:1 split

liner

4 mm ID FocusLiner with taper

sample

1. Fish Oil from capsule, prepared according to AOAC Method 991.39 2. Atlantic, farm-raised salmon, prepared according to AOCS Official Method Ce 1k-09

Descrição

Nota de análise

Samples of farm-raised Atlantic salmon were prepared and analyzed for fatty acids. The salmon was extracted using an acid digestion /alkali hydrolysis followed by methylation as described in AOCS Official Method Ce 1k-09 (2). The resulting extracts were then analyzed by GC/FID on an Omegawax 250 capillary column as described in AOCS Official Method Ce 1i-07.

Informações legais

null