Direkt zum Inhalt
Merck

TLC Analysis Of Plant Extracts on HPTLC Silica gel 60 NH2 F254s

TLC Analysis Of Plant Extracts on HPTLC Silica gel 60 NH2 F254s application for TLC

Materialien

Zubehör

Produkt-Nr.
Beschreibung
Preisangaben

Flaschenzerstäuber

capacity 240 mL

Rechteckige TLC-Entwicklungstanks, komplett

L × H × W 27.0 cm × 26.5 cm × 7.0 cm

Sigma TLC-Sprühkammer

Spectroline® UV-Lampe, E-Serie

output8 W, AC/DC input 115 V AC

Rechteckige TLC-Entwicklungstanks, komplett

L × H × W 7.5 cm × 15.5 cm × 8.0 cm

Hamilton® TLC-Spritzen

1702, volume 25 μL, needle size 22s ga, needle L 51 mm (2 in.),

coated tip

Spectroline® UV-Lampe, E-Serie

output6 W, AC/DC input 120 V AC

Hamilton® TLC-Spritzen

1705, volume 50 μL, needle size 22 ga, needle L 51 mm (2 in.),

coated tip

Hamilton® TLC-Spritzen

1701, volume 10 μL, needle size 26s ga, needle type, fixed, needle L 51 mm (2 in.),

coated tip

Rechteckige TLC-Entwicklungstanks, komplett

L × H × W 17.5 cm × 16.0 cm × 8.2 cm

Spectroline® UV-Lampe, E-Serie

output6 W, AC/DC input 230 V AC

Spectroline® UV-Lampe, E-Serie

output4 W, AC/DC input 230 V AC

Spectroline® UV-Lampe, E-Serie

output8 W, AC/DC input 230 V AC

Latch-Lid-Entwicklungskammer

For use with 10 cm × 10 cm plates

Hamilton® TLC-Spritzen

1710, volume 100 μL, needle size 22 ga, needle L 51 mm (2 in.),

coated tip

Latch-Lid-Entwicklungskammer

For use with 20 cm × 20 cm plates

Spectroline® UV-Lampe, E-Serie

output4 W, AC/DC input 115 V AC

Kimble® Mikrokapillarpipetten

Kimble® Mikrokapillarpipetten

Kimble® Mikrokapillarpipetten

capacity 20 μL, accuracy: ±0.5%

Reagenssprüher für DC-Platten

volume 250 mL, Bottle-Type

CONDITIONS

sample preparation

2 g plant material extracted for 1 h at 60°C under shaking with 50 ml solvent. Filtration of crude extract with an 0.45 μm syringe filter. Application by CAMAG ATS4 6 mm bandwise.

plate

1.15647 HPTLC Silica gel 60 RP-18 F 254s MS-grade 20cm x 10 cm (1.151647.0001)

eluent

[A] ethylacetat; [B] formic acid; [C] glacial acid; [D] water; (100:11:11:27, A:B:C:D) containing 0.1% TFA

migration distance

6 cm

chamber

normal chamber without chamber saturation

sample

0.150 μL

detector

UV, 254 nm (Identification by coupling the TLC – MS Interface (CAMAG) with the Advion expression) (CMS mass spectrometer (+)-ESI mode. Eluent: Acetonitrile / Water 95:5 + 0.1% TFA.)