capacity 32 oz (1000 mL), amber, white polypropylene cap, PTFE flat linerCONDITIONS
column
Equity-1701, 30 m x 0.25 mm I.D., 0.25 μm (28372-U)
oven
75 °C (2 min), 15 °C/min to 280 °C (5 min)
inj. temp.
250 °C
detector
ECD, 300 °C
carrier gas
helium, 1.2 mL/min. constant
injection
1 μL, splitless (splitter open after 0.75 min)
liner
4 mm I.D., single taper
sample
mixture of 608.1 and 608.2 calibration solutions (40351-U and 40352-U) diluted to 100 ppb in n-hexane