Przejdź do zawartości
Merck

G005512

EN 14110: GC Analysis of Methanol Impurity in Biodiesel on Equity-1 after SPME using 85 μm Polyacrylate Fiber

suitable for GC, application for SPME

Zaloguj sięWyświetlanie cen organizacyjnych i kontraktowych


About This Item

metody

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

parametry testowe

sample/matrix: 2 g of B100 biodiesel containing 0.01% methanol and 0.0785% 2-propanol in a 4 mL vial
SPME fiber: PK3 SPME ASSY 85UM POLYACRYLATE FS 24GA MAN (WHITE) (57304)
extraction: headspace, 60 °C for 5 min
desorption process: 250 °C for 0.75 min
column: Equity-1, 30 m x 0.32 mm I.D., 1.0 μm (28057-U)
oven: 50 °C
inj. temp.: 250 °C
detector: FID, 200 °C
carrier gas: helium, 30 cm/sec
liner: 0.75 mm I.D., SPME type, straight design (unpacked)

przydatność

application for SPME

Zastosowanie

biofuels
environmental

Informacje prawne

This page may contain text that has been machine translated.

kolumna analityczna


Wybierz jedną z najnowszych wersji:

Certyfikaty analizy (CoA)

Lot/Batch Number

Przepraszamy, ale COA dla tego produktu nie jest aktualnie dostępny online.

Proszę o kontakt, jeśli potrzebna jest pomoc Obsługa Klienta

Masz już ten produkt?

Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.

Odwiedź Bibliotekę dokumentów

Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.

Skontaktuj się z zespołem ds. pomocy technicznej