Passa al contenuto
Merck
Tutte le immagini(1)

Documenti fondamentali

59101

Supelco

Column Blank Kit for HPLC

I.D. × O.D. × L 4.6 mm × 1/4 in. × 15 cm

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali

Scegli un formato

1 EA
114,00 €

114,00 €


Per informazioni sulla disponibilità, contatta il Servizio Clienti.

Richiedi un ordine bulk

Scegli un formato

Cambia visualizzazione
1 EA
114,00 €

About This Item

Codice UNSPSC:
41115700

114,00 €


Per informazioni sulla disponibilità, contatta il Servizio Clienti.

Richiedi un ordine bulk

Materiali

316L stainless steel

Diam. int. × diam. est. × lungh.

4.6 mm × 1/4 in. × 15 cm

Cerchi prodotti simili? Visita Guida al confronto tra prodotti

Descrizione generale

Each column blank kit includes a precision-bore polished 316 stainless steel column blank, two zero dead volume external reducing unions (internal unions with 10 mm I.D. blanks) with nuts and ferrules for 1/16 in. O.D. tubing, and two
2 μm frits.

Prodotti consigliati

Discover LiChropur reagents ideal for HPLC or LC-MS analysis

Scegli una delle versioni più recenti:

Certificati d'analisi (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documenti section.

Se ti serve aiuto, non esitare a contattarci Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Domande

1–5 di 5 domande  
  1. How do I clean the HPLC Column Blank, Product No. 59101, after use?

    1 risposta
    1. First, check the instrument or pump manual to make sure the system can be used with a solution of 6-10% nitric acid.  Once all packing has been carefully removed from the column and disposed of properly, flush the column with hot HPLC grade or distilled water to remove residual packing or salts that may have precipitated during use.  Then passivate the column with a solution of 6-10% nitric acid at a flow rate of approximately 4-5 mL per minute.  Flush thoroughly with HPLC grade or distilled water, then methanol.

      Utile?

  2. What is the Department of Transportation shipping information for this product?

    1 risposta
    1. Transportation information can be found in Section 14 of the product's (M)SDS.To access the shipping information for this material, use the link on the product detail page for the product.

      Utile?

  3. What is the pressure rating for 59101 Column Blank Kit for HPLC?

    1 risposta
    1. Under its intended use, the standard 15cm x 4.6mm HPLC column blank should go no higher than 5,000 psi.

      Utile?

  4. I need extra fittings for my 1/4 inch OD Column Blank Kit (4.6mm ID), Product 59101.  Which fitting should be purchased?

    1 risposta
    1. The Empty Column Blank (Product No. 59101) comes with two size fittings: one that is a part of the column and one that connects 1/16 in. tubing to the column.  The column inlet/outlet fitting is a reducing union from 1/4 in.-1/16 in., Product No. Z226564.  The nuts/ferrule combination used to connect 1/16 in. tubing to the inlet and outlet of the column is a 10-32 Valco fitting. The product number for a pack of 10 nuts is 22990-U, and for the matching ferrules, Product No. 22988.

      Utile?

  5. I need to replace the frit in my 1/4 inch OD Column Blank Kit (4.6mm ID), Product 59101.  Which frit should be purchased?

    1 risposta
    1. The Empty Column Blank (Product No. 59101) requires a frit for a ¼ in. column, which comes in pore sizes of 0.5μm and 2μm.  The product numbers are 59038 and 58264, respectively.  We generally recommend the 0.5μm frits for particle sizes less than 5μm, and the 2μm for larger particle sized packings.

      Utile?

Recensioni

Nessuna valutazione

Filtri attivi

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.