Pular para o conteúdo
Merck

LC/MS/MS of Digoxin and Digitoxin in Serum or Plasma on Titan C18 after SPE with HybridSPE®-Phospholipid, Minimization of Matrix Effects

LC/MS/MS of Digoxin and Digitoxin in Serum or Plasma on Titan C18 after SPE with HybridSPE®-Phospholipid, Minimization of Matrix Effects application for LC-MS, application for SPE

Materiais

padrão

Nº do produto
Descrição
Preços

Digoxin solution

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Digitoxin solution

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

tubo ou placa SPE

Nº do produto
Descrição
Preços

HybridSPE®-PLus 96-Well Plate

volume 2 mL, pk of 1 ea

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

rat plasma or human serum spiked at 0.05 ng/mL with digitoxin and digoxin

SPE well plate

HybridSPE-PLus 96-well plate, 50 mg/well (575659-U)

sample addition

100 μL spiked rat plasma followed by 300 μL 1% formic acid in acetonitrile

elution

apply vacuum at 10 in Hg for 4 min

column

Titan C18, 10 cm x 2.1 mm I.D., 1.9 μm particles (577124-U)

mobile phase

10 mM ammonium formate in methanol:water (80:20)

flow rate

0.2 mL/min

pressure

4550 psi (314 bar)

detector

MS, ESI(+), MRM 798.5/651.5 (digoxin), 782.5/635.5 or 243.2 (digtoxin)

injection

2 μL

Descrição

Nota de análise

The cardiac glycosides digitoxin and digoxin are widely prescribed for treating congestive heart failure. Immunoassay-based methods are not able to effectively differentiate these two compounds due to their similar chemical structures. This application demonstrates a rapid, selective and sensitive LC/MS/MS method for determination of digitoxin and digoxin in plasma with a simple sample cleanup technique using HybridSPE-PLus for simultaneous removal of proteins and phospholipids and a Titan C18 column for rapid and efficient UHPLC separation. The highest grade LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable quantification.

Informações legais

HybridSPE is a registered trademark of Merck KGaA, Darmstadt, Germany
null