LC/MS Analysis of Antiarrhythmic Drugs and Metabolites on Ascentis® Express HILIC

LC/MS Analysis of Antiarrhythmic Drugs and Metabolites on Ascentis® Express HILIC application for HPLC

Materiais

Padrão

Nº do produto
Descrição
Preços

Amiodarone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

N-Desethylamiodarone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

(±)-Flecainide solution

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Lidocaine solution

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

coluna analítica

Nº do produto
Descrição
Preços

Ascentis® Express HILIC, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

tubo ou placa SPE

Nº do produto
Descrição
Preços

HybridSPE®-Phospholipid

96-well Plate, bed wt. 50 mg, volume 2 mL, pk of 1

CONDITIONS

column

Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)

mobile phase

[A] 5 mM ammonium formate; [B] 5 mM ammonium formate in acetonitrile, 5:95 (A:B), adjusted to pH 7.0 with formic acid

flow rate

0.4 mL/min

pressure

1305 psi (90 bar)

column temp.

35 °C

detector

ESI+, 100-1000 m/z

injection

0.5 μL

sample

each compound, 300 ng/mL in 75:25 (1% formic acid acetonitrile:water)

Descrição

Nota de análise

The basic nature of these antiarrhythmic agents makes them ideal candidates for HILIC (hydrophilic interaction) chromatographic separation. The benefit of HILIC over traditional reversed-phase chromatography is two-fold for both sample introduction and analyte detection. First, the high acetonitrile concentration of HILIC mobile phases allows for direct analysis of precipitated plasma samples without the need for additional sample solvent exchange. Second, the high acetonitrile content provides increased analyte response in positive ESI MS detection. Of the various HILIC-mode columns tested, method development for this assay determined the Ascentis Express HILIC, 2.7 μm particles, provided the best chromatographic resolution of the antiarrhythmic drugs while maintaining high peak efficiency for enhanced detection levels.