Pular para o conteúdo
Merck

HPLC Analysis of Substituted Poly Aromatic Hydrocarbons using Ascentis® Express Biphenyl

HPLC Analysis of Substituted Poly Aromatic Hydrocarbons using Ascentis® Express Biphenyl application for HPLC

Materiais

Padrão

Nº do produto
Descrição
Preços

Uracil

≥99.0%

Acenaphthene

99%

coluna analítica

Nº do produto
Descrição
Preços

Ascentis® Express Biphenyl Column

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

CONDITIONS

column

Ascentis Express Biphenyl, 10 cm x 2.1 mm, 2.7 μm particles (64065-U)

mobile phase

[A] Water; [B] Methanol; (30:70 A:B)

flow rate

0.5 mL/min

pressure

6200 psi (427 bar)

column temp.

35 °C

detector

UV, 254 nm

injection

1 μL

sample

Uracil (10 μg/mL), Acenaphthenol (250 μg/mL), Acenaphthenone (20 μg/mL),
  Acenaphthene (500 μg/mL), Prepared in 70:30 A:B

Descrição

Nota de análise

This application demonstrates the suitability of Ascentis Express Biphenyl for the separation of substituted PAHs. It shows the suitability of this phase for separations of extended aromatics and their oxygenated counter parts. These compounds are potentional metabolites, used as synthetic intermediates, and in the study of dihydrodiol dehydrogenase activity in celluar systems.

Informações legais

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany