Pular para o conteúdo
Merck

US EPA Method 610: GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) on SLB®-5ms

US EPA Method 610: GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) on SLB®-5ms suitable for GC

Materiais

coluna analítica

Nº do produto
Descrição
Preços

SPB®-608 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

CONDITIONS

column

SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (24103-U)

oven

70 °C (0.2 min), 25 °C/min to 265 °C, 5 °C/min to 315 °C (2 min)

inj. temp.

250 °C

detector

FlD, 350 °C

carrier gas

helium, 1.3 mL/min, constant

injection

2.0 μL pulsed splitless (30 psi until 0.2 min)

liner

4 mm I.D. single taper

sample

1 μg/mL PAH standard in methylene chloride

Descrição

Nota de análise

US EPA Method 610 describes the analysis of polynuclear aromatic hydrocarbons (commonly referred to as PAHs or PNAs) by both HPLC and GC. The method designates the use of a packed GC column, however capillary columns will provide better resolution, baseline stability, and longer lifetime. This application illustrates the separation of the target compound list in this method on the SLB-5ms capillary GC column. A pressure pulsed injection was used to improve response of the heavier PAHs, and to reduce .the larger vapor volume resulting from a 2 μL injection.

Informações legais

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany