Pular para o conteúdo
Merck

GC Analysis of Class 2 Residual Solvents on VOCOL® after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Class 2 Residual Solvents on VOCOL® after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

Materiais

Fibra para SPME

Nº do produto
Descrição
Preços

Conjunto de fibras para microextração em fase sólida (SPME) de poliacrilato (PA)

df 85 μm(PA, for use with manual holder, needle size 24 ga

CONDITIONS

sample/matrix

2 mL water + sodium chloride at 25% (pH 11) containing each analyte at 5 ppm, in a 4 mL vial

SPME fiber

85 μm Polyacrylate (57304)

extraction

heated headspace, 60 °C for 5 min

desorption process

250 °C for 3 min

column

VOCOL, 10 m x 0.20 mm I.D., 1.2 μm (24129-U)

oven

40 °C (0.75 min), 20 °C/min to 200 °C (10 min)

inj. temp.

250 °C

detector

FID

carrier gas

helium, 35 cm/sec @ 40 °C

injection

5:1 split

liner

0.75 mm I.D. SPME type

Descrição

Informações legais

VOCOL is a registered trademark of Merck KGaA, Darmstadt, Germany