LC/MS Analysis of Polypeptide Drugs on Ascentis® Express RP-Amide
Materiały
kolumna analityczna
Numer produktu
Opis
Cennik
składnik fazy ruchomej
Numer produktu
Opis
Cennik
CONDITIONS
column
Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2.7 μm particles (53913-U)
mobile phase
[A] 10 mM ammonium acetate in water, pH 4.2; [B] 10 mM ammonium acetate in 90:10 (v/v) acetonitrile:water, pH 4.7
gradient
held at 20% B for 0.1 min; 20 to 80% B in 2.9 min; 80 to 20% B in 0.01 min; held at 20% B for 0.99 min
flow rate
0.3 mL/min
column temp.
35 °C
detector
TOF, ESI(+), m/z 100-2500
injection
3 μL
sample
1 μL/mL in 10mM ammonium formate, 80:20, acetonitrile:water
Opis
Komentarz do analizy
This application demonstrates the suitability of the Ascentis Express RP Amide for the analysis of polypeptide drugs.
Informacje prawne
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany