Przejdź do zawartości
Merck

HPLC Analysis of Methylimidazoles in Diet Cola on Ascentis® Express HILIC

HPLC Analysis of Methylimidazoles in Diet Cola on Ascentis® Express HILIC application for HPLC

Materiały

standardowe

Numer produktu
Opis
Cennik

2-Methylimidazole

≥95.0% (HPLC), pharmaceutical impurity standard

CONDITIONS

column

Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)

mobile phase

2 mM ammonium formate in 5:95, water:acetonitrile, to pH 4.4 with formic acid

flow rate

0.6 mL/min

column temp.

50 °C

detector

ESI(+), 50-800 m/z

injection

1 μL

sample

200 mg/L in methanol

Opis

Opis ogólny

This method features fast analysis of methylimidazole derivatives used in caramel-colored carbonated beverages.

Komentarz do analizy

The purpose of this study was to develop a fast analytical method for analyzing for 2-methylimidazole and 4-methylimidazole in caramel colored carbonated beverages. The FDA has recently increased the awareness of the potential hazards of methylimidazole. These compounds can be found in caramel additives used in a broad range of food stuff. In this study the focus is the levels of methylimidazole in various colas. Methylimidazoles are a difficult group of compounds to analyze due their polar nature and low molecular weight. Traditional reversed phase separation are unsuccessful in retaining these small polar compounds, with most methods utilizing ion-exchange resins for analysis. Other common methods are GC analysis with sample derivitization.

Informacje prawne

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany