Przejdź do zawartości
Merck

HPLC Analysis of Sugars on Ascentis® Express HILIC (Mobile Phase 15:85)

HPLC Analysis of Sugars on Ascentis® Express HILIC (Mobile Phase 15:85) application for HPLC

Materiały

kolumna analityczna

Numer produktu
Opis
Cennik

Ascentis® Express HILIC, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 15 cm × 4.6 mm

standardowe

Numer produktu
Opis
Cennik

Maltotriose

≥90% (HPLC)

Sucrose

BioUltra, for molecular biology, ≥99.5% (HPLC)

Lactose Monohydrate

analytical standard

Glucose

European Pharmacopoeia (EP) Reference Standard

Xylose

European Pharmacopoeia (EP) Reference Standard

Maltose monohydrate

United States Pharmacopeia (USP) Reference Standard

CONDITIONS

column

Ascentis Express HILlC, 15 cm x 4.6 mm I.D., 2.7 μm particles (53981-U)

mobile phase

[A] water; [B] acetonitrile; (15:85, A:B)

flow rate

1.0 mL/min

column temp.

35 °C

detector

ELSD, 55 °C, 3.5 bar nitrogen

injection

10 μL

Opis

Komentarz do analizy

This application demonstrates the suitability of the Ascentis Express HILIC for the analysis of sugars. Structures along with the optimized chromatogram obtained on the Ascentis Express HILIC are presented. Separation is by normal-phase mechanism as a decrease in mobile phase water content results in increased retention.

Informacje prawne

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany