HPLC Analysis of Lysozyme on SRT® SEC-300, 30 cm x 7.8 mm I.D., 5 μm particles

application for HPLC

Zaloguj sięWyświetlanie cen organizacyjnych i kontraktowych


About This Item

metody

HPLC: suitable

parametry testowe

column: SRT SEC-300, 30 cm x 7.8 mm I.D., 5 μm particles (Z777051)
mobile phase: 50 mM phosphate buffered saline, 200 mM Na2SO4, pH 7.0
flow rate: 1 mL/min
column temp.: ambient
detector: UV, 214 nm
injection: 5 μL
sample: 1 mg/mL in 50 mM phosphate buffered saline

przydatność

application for HPLC

Zastosowanie

life science and biopharma

Komentarz do analizy

The positively-charged protein, lyzoyme (pI=11), exhibits excellent peak shape and recovery on SRT SEC-300 columns due to its highly-inert surface. Low protein recovery is usually attributed to 1) hydrophobic interactions with the stationary phase and 2) electrostatic interactions with the silica suface. SRT SEC-300 surface is hydrophilic and densely bonded to silica, therefore providing high mass recovery and good peak shape.

Informacje prawne

SRT is a registered trademark of Sepax Technologies
This page may contain text that has been machine translated.

Analit

Opis

    Lysozyme from chicken egg white, powder or granules, ≥90 %, ≥39,000 units/mg protein

Choose from one of the most recent versions:

Certyfikaty analizy (CoA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Obsługa Klienta

Masz już ten produkt?

Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.

Odwiedź Bibliotekę dokumentów

Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.

Skontaktuj się z zespołem ds. pomocy technicznej