G005879

LC/MS/MS Analysis of Epinephrine, Metanephrine and Metabolites on Ascentis® Express OH5

application for HPLC

Přihlásitk zobrazení cen stanovených pro organizaci a smluvních cen


About This Item

technique(s)

HPLC: suitable

test parameters

column: Ascentis Express OH5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53757-U)
mobile phase: [A] 5 mM ammonium formate; [B] 5 mM ammonium  formate in acetonitrile; (15:85, A:B)
flow rate: 0.3 mL/min
pressure: 1189 psi (82 bar)
column temp.: 35 °C
detector: MS ESI(+), 100-1000 m/z
injection: 1 μL
sample: 300 ng/mL each in acetonitrile

suitability

application for HPLC

application(s)

clinical

Analysis Note

HILIC mode on Ascentis Express OH5 provided rapid separation of neurotransmitters in highly MS-compatible mobile phases. The highest grade LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Vyberte jednu z posledních verzí:

Osvědčení o analýze (COA)

Lot/Batch Number

Je nám líto, ale pro tento produkt momentálně nemáme COA k dispozici online.

Potřebujete-li pomoc, obraťte se na Zákaznická podpora

Již tento produkt vlastníte?

Dokumenty související s produkty, které jste v minulosti zakoupili, byly za účelem usnadnění shromážděny ve vaší Knihovně dokumentů.

Navštívit knihovnu dokumentů

Náš tým vědeckých pracovníků má zkušenosti ve všech oblastech výzkumu, včetně přírodních věd, materiálových věd, chemické syntézy, chromatografie, analytiky a mnoha dalších..

Obraťte se na technický servis.