G003781
Fraction volatile du limoncello
application for SPME, suitable for GC
About This Item
Technique(s)
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
Paramètres du test
sample/matrix: Limoncello diluted 1:100 in water, then 5 mL of this solution transferred to a 10 mL crimped vial spiked with 100 μL of a 100 ppm solution of nonane (internal standard)
SPME fiber: 100 μm polydimethylsiloxane (57300-U)
extraction: headspace, with a pre-equilibration period of 10 min. at 40 °C, and then a fiber exposure of 30 min. at 40 °C, agitated in an alternate clockwise-anticlockwise rotation mode at 500 ppm
desorption process: 250 °C for 5 min
column: SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven: 40 °C, 3 °C/min to 250 °C (2 min)
inj. temp.: 250 °C
detector: FID, 260 °C (sampling rate 80 ms)
carrier gas: helium, 35.0 cm/sec
injection: splitless
liner: 0.75 mm I.D. SPME (2633901)
sample: Limoncello liquor
Adéquation
application for SPME
Application(s)
food and beverages
Remarque sur l'analyse
Informations légales
Analyte
Colonne analytique
Fibre SPME
Produit(s) apparenté(s)
Étalon
Choose from one of the most recent versions:
Certificats d'analyse (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique