跳轉至內容
Merck
全部照片(2)

重要文件

Z742475

BeadBlaster 24 Microtube homogenizer with 24 x 2.0 mL tube holder

AC/DC input 230 V AC, EuroPlug

同義詞:

bead mill, laboratory blender, laboratory homogenization, laboratory tube mill, laboratory tube mixer, tissue disruption

登入查看組織和合約定價


About This Item

分類程式碼代碼:
41121800
NACRES:
NB.11
暫時無法取得訂價和供貨情況

描述

cycle time 1 to 90 sec. (up to 90 sec.)

特點

program memory: up to 50
< 68 dB(A) Noise level

包裝

pkg of 1

製造商/商標名

Benchmark Scientific D2400-E

參數

(Pause between cycles: 0 to 2 min (1sec increment))
speed (4.0 to 7.0 m/s (0.05 m/s increments))

技術

PCR: suitable

尺寸

11 in. (28 cm) × 14.2 in. (36 cm) × 15.2 in. (38.5 cm)

重量

55 lb (25 kg)

容量

24 x 2.0 ml capacity

尋找類似的產品? 前往 產品比較指南

一般說明

The BeadBlaster 24 completely lyses, grinds and homogenizes a broad variety of biological samples. Even the most difficult and resistant samples are fully homogenized, often in 35 seconds or less.
  • For all sample types, from soft tissue to bone
  • Mix up to 24 tubes simultaneously
  • Sealed 2.0ml tubes - eliminates cross contamination

PLEASE ENSURE TUBES WITH 6mm HIGH CAPS ARE USED. 9mm HIGH CAP TUBES WILL STRESS THE TUBE HOLDER.

特點和優勢

  • Available as a full range of beads in prefilled tubes and bulk.
  • Digitally set and displayed operating parameters.
  • Transaparent lid for sample visualization.
  • The mechanical lid prevents accidental lid opening while operating.
  • Brushless motor lowers motor noise and ensures maintenance-free operation.

法律資訊

BeadBlaster is a trademark of Benchmark Scientific, Inc.

從最近期的版本中選擇一個:

分析證明 (COA)

Lot/Batch Number

抱歉,我們目前沒有在線上提供本產品的COA

如果您需要協助,請聯絡 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

Questions

  1. Does this homogenizer denature tissue proteins? For example, will homogenized tissues still be suitable for use in determining how small molecules are bound to the tissue proteins (either through a dialysis or ultracentrifugation assay)?

    1 answer
    1. This product is manufactured by Benchmark Scientific Inc. According to the supplier, proteins will generally survive the bead-beating process. It is recommended to monitor heat build-up. Start with all components cold and run the unit in short bursts with a rest between cycles. Don't overprocess the samples. If these guidelines are followed, multi-subunit enzymes can retain activity following bead beating. However, if the protein is membrane-bound, disruption may cause denaturation.

      For more information on this product, please view the manufacture website and data sheet below:
      https://www.benchmarkscientific.com/product/d2400/
      https://www.benchmarkscientific.com/wp-content/uploads/2020/10/BeadBlaster-24-jpgs-literature-rpm.pdf

      Helpful?

Reviews

No rating value

Active Filters

我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.

聯絡技術服務