Przejdź do zawartości
Merck

LC/MS Analysis of Peptides on Discovery® BIO Wide Pore C18, Effect of TFA Concentration on MS Signal

LC/MS Analysis of Peptides on Discovery® BIO Wide Pore C18, Effect of TFA Concentration on MS Signal application for HPLC

Materiały

kolumna analityczna

Numer produktu
Opis
Cennik

Discovery® BIO Wide Pore C18 HPLC Column

3 μm particle size, L × I.D. 15 cm × 2.1 mm

CONDITIONS

column

Discovery® BIO Wide Pore C18, 15 cm x 2.1 mm I.D., 3 μm particles (567202-U)

mobile phase

[A]: aqueous component (25 mM formic acid, 0.01% TFA or 0.1% TFA); [B] (50:50) water:acetonitrile containing same ionic additives as aq. component

flow rate

0.20 mL/min

column temp.

ambient

detector

(+)ESI

injection

1-3 μL

sample

RP Peptide Performance Standard (Alberta Peptide Inst., Cat. No. RPS-P0010)

Opis

Informacje prawne

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany