GC Analysis of PCB Congeners in Stream Sediment on SPB®-Octyl after SPME using 100 μm PDMS Fiber

GC Analysis of PCB Congeners in Stream Sediment on SPB®-Octyl after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

Materiały

Włókno SPME

Numer produktu
Opis
Cennik

SPME fiber assembly Polydimethylsiloxane (PDMS)

df 100 μm(PDMS, needle size 24 ga, for use with manual holder

kolumna analityczna

Numer produktu
Opis
Cennik

SPB®-Octyl Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

CONDITIONS

sample/matrix

5 g stream sediment, downstream from industrial sites

SPME fiber

PDMS, 100 μm (57300-U) (57300-U)

extraction

headspace, 90 °C/30 min

desorption process

280 °C/2 min, splitless

column

SPB®-Octyl, 30 m × 0.25 mm I.D., 0.25 μm (24218-U)

oven

50 °C/1 min at 10 °C/min to 150 °C/0 min