GC Analysis of Hydrocarbons in Gasoline on Petrocol® DH, Isothermal
Materiały
kolumna analityczna
Numer produktu
Opis
Cennik
CONDITIONS
column
Petrocol DH, 100 m x 0.25 mm I.D., 0.50 μm (24160-U)
oven
60 °C
inj. temp.
250 °C
detector
FID, 250 ºC
carrier gas
helium, 19-21 cm/sec
injection
0.1 μL, 100:1 split
Opis
Komentarz do analizy
Detailed hydrocarbon analysis (DHA) is performed on long non-polar columns. Elution occurs in order of lowest boiling point to highest. The chromatogram shown here is of a gasoline sample using a 100 m Petrocol DH column. The long column length and 60 ºC isothermal oven temperature allow the separation of many components in this complex sample. Compounds identified range from propane (peak 1) to 1-methylnaphthalene (peak 96). Retention index libraries exist for this column, adding to its usefulness.
Informacje prawne
Petrocol is a registered trademark of Merck KGaA, Darmstadt, Germany