Przejdź do zawartości
Merck

GC Analysis of Cocaine in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber

GC Analysis of Cocaine in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

Materiały

Włókno SPME

Numer produktu
Opis
Cennik

SPME Fiber Assembly

100 μm PDMS, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, plain red hub

kolumna analityczna

Numer produktu
Opis
Cennik

Equity-1 Capillary GC Column

L × I.D. 30 m × 0.32 mm, df 0.25 μm

CONDITIONS

sample/matrix

0.5 mL urine (250 ng each analyte, 20 μL 2.5% NaF) in 1 mL vial

SPME fiber

polydimethylsiloxane, 100 μm (57300-U)

extraction

immersion, 30 min

desorption process

3 min, 240 °C

column

100% methyl polysiloxane, 30 m × 0.32 mm I.D., 0.25 μm (Supelco equivalent, Equity-1, 28055-U)

oven

120 °C to 280 °C at 10 °C/min

inj. temp.

splitless (splitter opened after 1 min), 240 °C

detector

NPD, 280 °C

carrier gas

helium, 3 mL/min