GC Analysis of Class 3 Residual Solvents on Equity-1 after SPME using 100 μm PDMS Fiber

GC Analysis of Class 3 Residual Solvents on Equity-1 after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

Materiały

Włókno SPME

Numer produktu
Opis
Cennik

SPME fiber assembly Polydimethylsiloxane (PDMS)

df 100 μm(PDMS, needle size 24 ga, for use with manual holder

kolumna analityczna

Numer produktu
Opis
Cennik

Equity-1 Capillary GC Column

L × I.D. 10 m × 0.20 mm, df 1.20 μm

CONDITIONS

sample/matrix

2 mL water + sodium chloride at 25% (pH 2) containing each analyte at 5 ppm, in a 4 mL vial

SPME fiber

100 μm PDMS, 24 ga manual (57300-U)

extraction

heated headspace, 60 °C for 5 min

desorption process

250 °C for 3 min

column

Equity-1, 10 m x 0.20 mm I.D., 1.2 μm (28043-U)

oven

40 °C (0.75 min), 20 °C/min to 200 °C (10 min)

inj. temp.

250 °C

detector

FID

carrier gas

helium, 35 cm/sec @ 4

injection

5:1 split

liner

0.75 mm I.D. SPME type

Opis

Informacje prawne