96-0114
GC Analysis of Residual Solvents on SPB®-1 SULFUR after SPME using 65 μm PDMS/DVB Fiber
suitable for GC, application for SPME
About This Item
Polecane produkty
metody
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
parametry testowe
sample/matrix: water + 25% NaCl, pH 11.0 immersion: 3.9 mL sample + 0.1 mL 50% Na3PO4 / 1 g NaCl headspace: 2.4 mL sample + 0.06 mL 50% Na3PO4 / 0.6 g NaCl
SPME fiber: polydimethylsiloxane/DVB, 65 μm (57310-U)
extraction: immersion, 15 min (rapid stirring or headspace, 15 min, 55 °C)
desorption process: 270 °C, 5 min
column: SPB-1 SULFUR, 30 m × 0.32 mm I.D., 4.0 μm (24158)
oven: 50 °C (2 min) to 180 °C at 10 °C/min
inj. temp.: splitless (3 min), 270 °C (0.75 mm I.D. liner)
detector: FID
carrier gas: helium, 50 cm/sec
przydatność
application for SPME
Zastosowanie
pharmaceutical (small molecule)
Informacje prawne
Analit
Włókno SPME
kolumna analityczna
Choose from one of the most recent versions:
Certyfikaty analizy (CoA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Obsługa Klienta
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej