コンテンツへスキップ
Merck

LC/MS Analysis of Illicit Phenethylamine and Cathinone “Bath Salts” in Plasma on Ascentis® Express HILIC after Extraction using SPME LC Tips

LC/MS Analysis of Illicit Phenethylamine and Cathinone “Bath Salts” in Plasma on Ascentis® Express HILIC after Extraction using SPME LC Tips application for LC-MS, application for SPME

材料

Standard

製品番号
詳細
価格

Ethylone hydrochloride

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Mephedrone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Methylone hydrochloride

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

付属品

製品番号
詳細
価格

IKA® MS 3 digital shakers

universal small shaker, AC/DC input 115 V AC, US 2-pin plug

分析カラム

製品番号
詳細
価格

Ascentis® Express HILIC、2.7 μm HPLCカラム

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

移動相成分

製品番号
詳細
価格

ギ酸アンモニウム

eluent additive for LC-MS, LiChropur, ≥99.0%

CONDITIONS

sample/matrix

Rat plasma was adjusted to pH 3.0 with phosphoric acid and spiked at 50 ng/mL with selected Bath Salt compounds. (A 750 μL aliquot of the spiked plasma was placed into each well of a 1.0 mL 96 well collection plate.)

SPME fiber

SPME LC tip, mixed-mode (57234-U)

extraction

SPME LC fibers were preconditioned in 50:50 methanol:water and placed into sample wells. Extraction was performed for 15 minutes on an IKA VORTEX MS3 shaker at 500 rpm.

desorption process

SPME LC Tip fibers were then desorbed in 100 μL of 0.5% ammonium hydroxide in methanol for 60 minutes at 500 rpm vortex.

desorption process

Desorbed samples were evaporated and reconstituted in 40 μL of acetonitrile and injected directly into the LC/MS system.

column

Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)

mobile phase

5 mM ammonium formate (98:2 acetonitrile:water)

flow rate

0.6 mL/min

pressure

1842 psi (127 bar)

column temp.

35 °C

detector

ESI(+), 100-1000 m/z

injection

1 μL

詳細

アナリシスノート

Solid phase microextraction (SPME) was used in this application to extract phenethylamine and cathinone "bath salts" from spiked plasma. After elution from the SPME fiber, the compounds were separated on an Ascentis Express HILIC column and detected using MS ESI(+) detection. Ultra high purity LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.

法的情報

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany