LC/MS/MS Analysis of Pesticides in Cannabis on the Ascentis® RP-Amide after QuEChERS Extraction and Cleanup Using Supel QuE Verde

LC/MS/MS Analysis of Pesticides in Cannabis on the Ascentis® RP-Amide after QuEChERS Extraction and Cleanup Using Supel™ QuE Verde application for LC-MS, application for SPE

Materiali

Accessorio

N° Catalogo
Descrizione
Determinazione del prezzo

Certified Vial Kit, Low Adsorption (LA), 2 mL, pk of 100

volume 2 mL, amber glass vial (with marking spot), natural PTFE/silicone septa (with slit), thread for 9 mm

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

Ascentis® RP-Amide HPLC Column

3 μm particle size, L × I.D. 10 cm × 2.1 mm

Preparazione del campione

N° Catalogo
Descrizione
Determinazione del prezzo

QuEChERS Shaker and Rack Starter Kit

AC/DC input 115 V AC (USA Compatible Plug)

Solvente

N° Catalogo
Descrizione
Determinazione del prezzo

Acetonitrile

HPLC Plus, ≥99.9%

CONDITIONS

sample preparation

Dispersive (QuEChERS)

sample/matrix

1.9 g of cannabis spiked with pesticides at 50 ng/g

extraction tube

Supel QuE Citrate extraction tube (55227-U)

extraction process

add 10 mL of acetonitrile; shake for 30 minutes at 2500 rpm on shaker; add contents of Supel QuE citrate extraction tube; (shake immediately for 1 minute; centrifuge at 5000 rpm for 5 minutes)

clean-up tube

Supel QuE Verde, 2 mL (55227-U)

clean-up process

transfer 1 mL of acetonitrile layer into Supel QuE Verde cleanup tube; shake for 1 minutes; (centrifuge at 5000 rpm for 3 minutes; draw of supernatant for LC/MS/MS analysis)

column

Ascentis RP-Amide, 10 cm x 2.1 mm I.D., 3.0 μm particles (565301-U)

column

Ascentis RP-Amide Supelguard guard cartridge, 2 cm x 2.1 mm I.D., 5 μm particles (565372-U)

mobile phase

[A] 5 mM ammonium formate, 0.1% formic acid in 95:5 water:acetonitrile; [B] 5 mM ammonium formate, 0.1% formic acid in 5:95 water:acetonitrile

gradient

10% B held for 1 min; to 100% B in 13 min; held at 100% B for 6 min; to 10% in 0.5 min; held at 10% B for 6 min

flow rate

0.4 mL/min

column temp.

30 °C

detector

MRM

injection

5 μL

sample

QuEChERS extract of cannabis in acetonitrile

Descrizione

Risultati analitici

Consumption of cannabis and/or cannabis based products is currently legal in some form in 28 US states plus the District of Columbia. Pesticide testing on plant material and extracts is required in some states, including Oregon, which mandates a specific target list for testing. A common approach used by testing labs for pesticides is QuEChERS extraction followed by a dispersive cleanup. In this application a majority of the LC-amenable pesticides on the Oregon list were extracted from cannabis spiked at 50 ng/g using QuEChERS, followed by cleanup with Supel QuE Verde. Analysis was achieved using the Ascentis® RP-Amide column equipped with a guard. Supel QuE Verde offered sufficient background reduction for LC/MS/MS analysis with recoveries of 70-120% for a majority of the pesticides.

Altre note

Sample pretreatment: Added 10 mL of deionized water, mixed and equilibrated for 10 minutes