713-0854
GC Analysis of PCBs in Transformer Oil on SPB®-5 after SPE using Supelclean™ LC-Florisil® and LC-Si
suitable for GC, application for SPE
About This Item
Prodotti consigliati
tecniche
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
Parametri analitici
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: used transormer oil containing 50 ppb Aroclor 1254.
SPE tube/cartridge: Supelclean LC-Florisil, 1.0 g/6 mL (57057)
SPE tube/cartridge: Supelclean LC-Si, 1.0 g/6 mL (57051)
condition: 0.5 mL isooctane (Florisil tube)
sample addition: 0.2 g transformer oil or 0.1 g waste oil
washing: 5 x 2 mL isooctane
column: SPB-5, 30 m × 0.32 mm I.D., 0.25 μm film (24048)
oven: 40 °C (4 min) to 300 °C at 10 °C/min, hold 5 min
detector: ECD
carrier gas: nitrogen
injection: 1 μL, splitless (30 sec delay), then split (50:1)