跳转至内容
Merck

USP HPLC Organic Impurity Analysis for Lamotrigine Tablets (SST) on an Ascentis® Express C18

USP HPLC Organic Impurity Analysis for Lamotrigine Tablets (SST) on an Ascentis® Express C18

材料

分析色谱柱

产品编号
说明
价格

Ascentis® Express C18,5μm HPLC色谱柱

5 μm particle size, L × I.D. 25 cm × 4.6 mm

标准

产品编号
说明
价格

拉莫三嗪

United States Pharmacopeia (USP) Reference Standard

2,3-二氯苯甲酸

United States Pharmacopeia (USP) Reference Standard

样品制备

产品编号
说明
价格

Millex PVDF 针式过滤器

pore size 0.45 μm, diam. 33 mm, non-sterile, hydrophilic

流动相组分

产品编号
说明
价格

Lamotrigine Related Compound C

United States Pharmacopeia (USP) Reference Standard

乙腈

isocratic grade for liquid chromatography LiChrosolv®

甲醇

for liquid chromatography LiChrosolv®

乙酸铵

for HPLC LiChropur

for chromatography (LC-MS Grade) LiChrosolv®

CONDITIONS

column

Ascentis® Express C18, 250 x 4.6 mm I.D., 5 µm (50538-U)

mobile phase

Acetonitrile:methanol:buffer (10:30:60) v/v/v; Buffer– 0.8 g/L of ammonium acetate, adjusted with glacial acetic acid to a pH of 4.5

gradient

Isocratic

flow rate

1 mL/min

pressure

216 bar (3132 psi)

column temp.

Ambient

detector

UV, 210 nm

injection

5 µL

sample/matrix

System suitability solution - 1.0 µg/mL of lamotrigine related compound B and 0.4 mg/ml of lamotrigine in methanol and buffer (60:40) v/v

说明

一般描述

Lamotrigine is an anti-convulsant medication used to treat epilepsy and to delay or prevent the recurrence of depressive episodes in bipolar disorder. For epilepsy, this includes focal seizures, tonic-clonic seizures, and seizures in Lennox-Gastaut syndrome.
In bipolar disorder, lamotrigine has not been shown to reliably treat acute depression, but for patients with bipolar disorder who are not currently symptomatic, it appears to be effective in reducing the risk of future episodes of depression.
This application illustrated the system suitability for the organic impurity analysis for lamotrigine tablets testing following the currently official United States Pharmacopoeia monograph

分析说明

For the lamotrigine standard solution, the RSD was 0.43 (USP specification NMT 10% RSD) and the tailing factor for lamotrigine was 1.7 (USP specification NMT 2). For the displayed system suitability solution the resolution between lamotrigine and the related compound B was 2.6 (USP specification NLT 2.0).

其他说明

App_367I

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany