HPLC Analysis of Sugars and Taurine in an Energy Drink on Ascentis® Express HILIC using UV and ELSD Detection in Series

application for HPLC

Přihlásitk zobrazení cen stanovených pro organizaci a smluvních cen


About This Item

technique(s)

HPLC: suitable

test parameters

column: Ascentis Express HILlC, 10 cm x 3.0 mm I.D., 2.7 μm particles (53970-U)
mobile phase: [A] 100 mM ammonium acetate, pH 5.0 with acetic acid; [B] water; [C] acetonitrile; (09:01:90, A:B:C)
flow rate: 0.6 mL/min
pressure: 812 psi (56 bar)
column temp.: 35 °C
detector: ELSD, 55 °C, 3.5 bar nitrogen
injection: 2 μL
sample: dilute1:9 in acetonitrile

suitability

application for HPLC

application(s)

food and beverages

Analysis Note

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of common drink additives used in caffeinated energy drinks. The analysis was performed with UV and ELSD detectors in series, this chromatogram being from the ELSD.

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyt

Popis

    Sucrose
    Taurine

analytical column

standard

Č. produktu
Popis
Stanovení ceny

Vyberte jednu z posledních verzí:

Osvědčení o analýze (COA)

Lot/Batch Number

Je nám líto, ale pro tento produkt momentálně nemáme COA k dispozici online.

Potřebujete-li pomoc, obraťte se na Zákaznická podpora

Již tento produkt vlastníte?

Dokumenty související s produkty, které jste v minulosti zakoupili, byly za účelem usnadnění shromážděny ve vaší Knihovně dokumentů.

Navštívit knihovnu dokumentů

Náš tým vědeckých pracovníků má zkušenosti ve všech oblastech výzkumu, včetně přírodních věd, materiálových věd, chemické syntézy, chromatografie, analytiky a mnoha dalších..

Obraťte se na technický servis.