G006299
LC/MS Analysis of Opioid Glucuronide Metabolites in Urine on Ascentis® Express F5 after Solid Phase Extraction (SPE) using Supel™-Select HLB
application for SPE, application for HPLC
About This Item
Methode(n)
HPLC: suitable
solid phase extraction (SPE): suitable
Testparameter
sample preparation: Solid Phase Extraction
sample/matrix: urine spiked with drugs and metabolites at 300 ng/mL
SPE well plate: Supel-Select HLB 96-well plate, 60 mg/well (575662-U)
condition: 1 mL methanol followed by 1 mL water
sample addition: 1 mL hydrolyzed urine
washing: 1 mL water followed by 1 mL 25% methanol
elution: 1 mL acetonitrile
column: Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)
mobile phase: [A] 5 mM ammonium formate, pH 4.2 (adjusted with formic acid); [B] 5 mM ammonium formate in (95:5 acetonitrile:water) pH 5.2 (adjusted with formic acid)
gradient: 15% B held for 1 min, to 70% B in 2 min, held at 70% B for 7 min
flow rate: 0.4 mL/min
column temp.: 35 °C
detector: ESI(+) TOF, full scan, m/z 100-1000
injection: 3 μL
sample: 300 ng/mL
Eignung
application for HPLC
application for SPE
Anwendung(en)
clinical
forensics and toxicology
Hinweis zur Analyse
Rechtliche Hinweise
Analyt
Analytische Säule
SPE-Röhrchen oder -Platte
Hier finden Sie alle aktuellen Versionen:
Analysenzertifikate (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Dokumente section.
Wenn Sie Hilfe benötigen, wenden Sie sich bitte an Kundensupport
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..
Setzen Sie sich mit dem technischen Dienst in Verbindung.