Direkt zum Inhalt
Merck

LC/MS Analysis of Opioid Glucuronide Metabolites in Urine on Ascentis® Express F5 after Solid Phase Extraction (SPE) using Supel-Select HLB

LC/MS Analysis of Opioid Glucuronide Metabolites in Urine on Ascentis® Express F5 after Solid Phase Extraction (SPE) using Supel™-Select HLB application for SPE, application for HPLC

Materialien

Analytische Säule

Produkt-Nr.
Beschreibung
Preisangaben

Ascentis® Express F5, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

SPE-Röhrchen oder -Platte

Produkt-Nr.
Beschreibung
Preisangaben

Supel Select HLB SPE 96-well Plate

bed wt. 60 mg/well, pk of 1

Standard

Produkt-Nr.
Beschreibung
Preisangaben

Morphine-3-β-D-glucuronide solution

1 mg/mL in methanol with 0.05% NaOH, ampule of 1 mL, certified reference material, Cerilliant®

Hydromorphon -Lösung

1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Buprenorphin-Glucuronid -Lösung

100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Buprenorphin -Lösung

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Kodein -Lösung

100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Codein-6-β-D-Glucuronid -Lösung

100 μg/mL in methanol: water (2:8), ampule of 1 mL, certified reference material, Cerilliant®

Hydrocodon -Lösung

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Hydromorphon-3-β-D-Glucuronid

100 μg/mL in methanol: water (1:1), ampule of 1 mL, certified reference material, Cerilliant®

Morphin -Lösung

100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Glucuronid-Norbuprenorphin -Lösung

100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Norbuprenorphin -Lösung

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Oxycodon -Lösung

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Oxymorphon-D3 -Lösung

100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Oxymorphon-3-β-D-Glucuronid -Lösung

100 μg/mL in methanol: water (1:1), ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

sample preparation

Solid Phase Extraction

sample/matrix

urine spiked with drugs and metabolites at 300 ng/mL

SPE well plate

Supel-Select HLB 96-well plate, 60 mg/well (575662-U)

condition

1 mL methanol followed by 1 mL water

sample addition

1 mL hydrolyzed urine

washing

1 mL water followed by 1 mL 25% methanol

elution

1 mL acetonitrile

column

Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)

mobile phase

[A] 5 mM ammonium formate, pH 4.2 (adjusted with formic acid); [B] 5 mM ammonium formate in (95:5 acetonitrile:water) pH 5.2 (adjusted with formic acid)

gradient

15% B held for 1 min, to 70% B in 2 min, held at 70% B for 7 min

flow rate

0.4 mL/min

column temp.

35 °C

detector

ESI(+) TOF, full scan, m/z 100-1000

injection

3 μL

sample

300 ng/mL

Beschreibung

Hinweis zur Analyse

Sample pretreatment: β-glucuronidase hydrolysis
The workflow for the LC/MS analysis of native drugs and associated metabolites from urine is shown here. The convenient liquid form of beta-glucuronidase from Sigma was used for hydrolysis of the glucuronide metabolites. SPE using Supel-Select HLB yielded clean extracts. The Ascentis Express F5 columns gave highly efficient, well-resolved peaks and rapid separation. Ultra high purity LC-MS solvents and additives provided trouble-free operation, and Cerilliant CRMs gave reliable quantification.

Rechtliche Hinweise

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC