跳轉至內容
Merck

HPLC Analysis of Protein Standards on Zenix® SEC-300 by Size Exclusion As Affected by Arginine

HPLC Analysis of Protein Standards on Zenix® SEC-300 by Size Exclusion As Affected by Arginine application for HPLC

材料

分析柱

產品號碼
描述
訂價

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 4.6 mm, 3 μm particle size, stainless steel

流動相成分

產品號碼
描述
訂價

L-精氨酸

from non-animal source, meets EP, USP testing specifications, suitable for cell culture, 98.5-101.0%

附屬品

產品號碼
描述
訂價

OPTI-SOLV EXP 在线过滤器

holder with EXP titanium hybrid ferrule, pack of 1 sets

EXP® Pre-Column Filter Cartridges

0.5μm, pkg of 5 ea

CONDITIONS

column

Zenix SEC-300, 30 cm x 4.6 mm I.D., 3 μm particles (A777028)

mobile phase

[A], 0.15 M potassium phosphate, monobasic, pH 7.0 with potassium hydroxide; [B] 0.15 M potassium phosphate, monobasic, 0.2 M arginine, pH 7.0 with phosphoric acid (mobile phase is either 100% A or 100% B)

flow rate

0.3 mL/min

pressure

1030 psi 100% A (71 bar)

pressure

1165 psi 100% B (80 bar)

column temp.

25 °C

detector

UV, 280 nm

injection

2.5 μL

sample

Waters BEH200 SEC protein standard

描述

分析報告

Arginine can have benefitial effects on peak shape and recovery of hydrophobic proteins when chromatographed by size exclusion chromatography. In this case, with protein standards that are not particularly hydrophobic, no dramatic affect is noted, as assessed by peak efficiency, peak widths, or peak retention.

法律資訊

Zenix is a registered trademark of Sepax Technologies