GC Analysis of Amphetamines in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber
材料
固相微萃取纖維
產品號碼
描述
訂價
CONDITIONS
sample/matrix
1 mL urine + 0.7g potassium carbonate in 20 mL headspace vial, equilibrated at 80°C, 30 min
SPME fiber
polydimethylsiloxane, 100 μm (57300-U)
derivatization
methyl bis-trifluoroacetamide (headspace, 0.5 min, ambient)
extraction
headspace, 3-5 min, 80 °C
desorption process
1 min, 270 °C
column
100% methyl polysiloxane, 12.5 m × 0.2 mm I.D., 0.33 μm (Supelco equivalent, Equity-1, 28041-U)
oven
60 °C (1 min) to 140 °C (4 min) at 30 °C/min, then to 276 °C at 20 °C/min, 4 min
inj. temp.
splitless (closed 1 min), 270 °C
detector
MS, full scan