HPLC: suitable
測試參數
column: Ascentis Express C18, 10 cm x 4.6 mm I.D., 5 μm particles (50536-U)
mobile phase: [A] water with 3% acetic acid; [B] methanol with 3% acetic acid; (79:21, A:B)
flow rate: 2.0 mL/min
pressure: 3290 psi (227 bar)
column temp.: 45 °C
detector: UV,275 nm
injection: 10 μL
sample: 100 - 360 μg/mLin mobile phase
適合性
application for HPLC
應用
clinical
pharmaceutical (small molecule)
分析報告
This application shows the improvements in efficiency that may be obtained for USP methods using modern Fused-Core technology.
法律資訊
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
分析物
描述
Acetaminophen, meets USP testing specifications, 98.0-102.0%, powder
Caffeine, meets USP testing specifications, anhydrous
Benzoic acid, meets USP testing specifications