跳轉至內容
Merck
全部照片(1)

重要文件

G005634

HPLC Analysis of Drugs of Abuse on Ascentis® Express F5

application for HPLC

登入查看組織和合約定價


About This Item

技術

HPLC: suitable

測試參數

column: Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)
mobile phase: [A] 5 mM ammonium formate; [B] 5 mM ammonium formate in 95:5 (v/v) acetonitrile:water
gradient: held at 20% B for 0.5 min; 20 to 100% B in 5.5 min; held at 100% B for 4 min
flow rate: 0.3 mL/min
pressure: 175 bar
column temp.: 35 °C
detector: ESI(+), m/z 100-1000
injection: 1 μL
sample: 1 μg/mL in 50:50, water: methanol

適合性

application for HPLC

應用

forensics and toxicology

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

描述

    Benzoylecgonine
    Morphine
    Codeine
    Oxycodone
    Hydrocodone
    (±)-Deoxyephedrine hydrochloride
    Fentanyl

從最近期的版本中選擇一個:

分析證明 (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our 文件 section.

如果您需要協助,請聯絡 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.

聯絡技術服務