To reduce the occurrence of sample-related bead aggregation, the following steps can be taken:
•Spin samples at 10,000 x g for 10 minutes immediately before preparing sample dilutions or loading samples into the plate.
•Resuspend the plate in Wash Buffer instead of Sheath or Drive Fluid. The plate must be read within a couple of hours, and the buffer should be switched to Sheath or Drive Fluid for longer storage.
•Dilute samples at a ratio of 1:2 or 1:4.
•Perform more wash cycles after the primary incubation step of the assay to remove substances that cause bead aggregation.
MCYTOMAG-70K-PMX
MILLIPLEX®小鼠细胞因子/趋化因子磁珠试剂盒 - 预混25重 - 免疫学多重分析
Simultaneously analyze multiple cytokine and chemokine biomarkers with Bead-Based Multiplex Assays using the Luminex technology, in mouse serum, plasma and cell culture samples.
同義詞:
Luminex® Mouse cytokine immunoassay, Millipore Mouse cytokine immunoassay, Mouse cytokine Multiplex Immunoassay
About This Item
推薦產品
品質等級
物種活性
mouse
製造商/商標名
Milliplex®
assay range
accuracy: 85-107%
standard curve range: 3.2-10,000 pg/mL
技術
multiplexing: suitable
相容性
configured for Premixed
檢測方法
fluorometric (Luminex xMAP)
運輸包裝
wet ice
一般說明
- 更柔韧的板和洗板机选择
- 使用浑浊的血清/血浆样品改善性能
- 测定结果等同于非微珠
- 自动洗涤避免了与真空过滤洗涤相关的许多问题
MILLIPLEX®小鼠细胞因子/趋化因子组套使您能够专注于细胞因子的治疗潜力以及细胞因子表达的调节。结合微珠格式的Luminex ® xMAP®平台,您可以获得理想的速度和灵敏度的优势,允许同时对数十种分析物进行定量多重检测,从而可以显著提高生产力。
组套类型:细胞因子/趋化因子
應用
- 分析物:G-CSF, GM-CSF, IFN-γ, IL-1α, IL-1β, IL-2, IL-4, IL-5, IL-6, IL-7, IL-9, IL-10, IL-12 (p40), IL-12 (p70), IL-13, IL-15, IL-17, IP-10, KC, MCP-1, MIP-1α, MIP-1β, MIP-2, RANTES, TNF-α
其他說明
法律資訊
訊號詞
Danger
危險分類
Acute Tox. 4 Dermal - Acute Tox. 4 Inhalation - Acute Tox. 4 Oral - Aquatic Chronic 2 - Eye Dam. 1 - Skin Sens. 1 - STOT RE 2
標靶器官
Respiratory Tract
儲存類別代碼
10 - Combustible liquids
分析證明 (COA)
輸入產品批次/批號來搜索 分析證明 (COA)。在產品’s標籤上找到批次和批號,寫有 ‘Lot’或‘Batch’.。
-
It seems that the sample wells are experiencing low bead counts while the standard and control wells are unaffected. This issue is limited to 1 or 2 cytokines, while the rest show normal counts. It doesn't appear to be related to the instrument, bead loss, or inadequate bead mixing. What could be the possible reason for this discrepancy?
1 answer-
Helpful?
-
-
Does the addition of 25 µL of Assay Buffer to the sample well count as a dilution when using a Milliplex kit that does not require a sample dilution prior to loading the plate?
1 answer-
If the sample is not diluted prior to loading the plate, and the same volume of standard and sample are loaded, then there is no need to apply a sample dilution factor.