參數

−5-110 °C temperature

尋找類似的產品? 前往 產品比較指南

一般說明

For applications where a separate reference electrode is required. Fast response over the full pH range with low sodium ion error. Useful for highly acidic or highly alkaline samples.

Help in selecting the correct electrode can be viewed at Electrode Selection Guide
Hints on Electrode Maintenance

法律資訊

Aldrich is a registered trademark of Sigma-Aldrich Co. LLC

相關產品


從最近期的版本中選擇一個:

分析證明 (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our 文件 section.

如果您需要協助,請聯絡 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

Questions

1–2 of 2 Questions  
  1. Is the electrode "Z264946" refillable?

    1 answer
    1. Yes, it can be re-filled.

      Helpful?

  2. Please provide user guide for Z264946 Aldrich® glass pH electrode.

    1 answer
    1. A user guide is not available for this product. See the usage instructions below:
      PREPARATION FOR USE:
      1. Wet lower end of electrode in tap water. Carefully remove the lower cap.
      NOTE: The cap is used only for shipping and storage. You might want to keep it for later use.
      2. If air bubbles are present in the bulb area, shake downwards to displace any air bubbles in the bulb.
      pH MEASUREMENT:
      1. Thoroughly rinse the electrode tip in distilled water and insert electrode tip in a known buffer, close to the sample to be tested, along with an appropriate reference electrode.
      2. Adjust the ‘standardize’ control on the pH meter until the meter reads the value of the buffer.
      3. Thoroughly rinse the electrodes with distilled water, then insert the electrodes in the solution to be tested.
      NOTE: The electrode tips should be rinsed in distilled water between each measurement to prevent contamination of the next solution.
      STORAGE PROCEDURES:
      1. For short storage (up to one week), the electrode may be placed in a beaker containing approximately 1 inch saturated KCl solution.
      2. For long-term storage (over one week), fill storage cap with KCl solution and place the cap back on the electrode tip.
      The electrode may then be placed back in the box and stored.