795-0298
EPA Method TO-11/IP-6A and ASTM® D5197: HPLC Analysis of Aldehydes and Ketones on SUPELCOSIL™ LC-18 after Collection/Desorption using LpDNPH
application for air monitoring, application for HPLC
About This Item
metody
HPLC: suitable
parametry testowe
sample preparation: SPE (Solid Phase Extraction)
adsorbent tube: LpDNPH Cartridge (21024-U)
sample/matrix: 5 μg each carbonyl-DNPH on LpDNPH cartridge. Cartridge eluted with 5.0 mL acetonitrile
column: SUPELCOSIL LC-18, 25 cm × 4.6 mm I.D., 5 μm particles (58298)
mobile phase: [A]:acetonitrile:tetrahydrofuran:water, 30:10:60; [B]: acetonitrile:water, 60:40
gradient: gradient 0% B for 1 min, linear gradient to 100% B over 10 min
flow rate: 1.5 mL/min
detector: VIS, 360 nm
injection: 25 μL of extract
przydatność
application for HPLC
application for air monitoring
Zastosowanie
air monitoring
industrial hygiene
Informacje prawne
Analit
kolumna analityczna
produkt powiązany
Choose from one of the most recent versions:
Certyfikaty analizy (CoA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Obsługa Klienta
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej