GC Analysis of a 20-Component Organochlorine Pesticide Mix in Water on a 5% Phenyl Column after SPE using Supelclean™ ENVI™-8
suitable for GC, application for SPE
Synonim(y):
Chlorinated pesticides in hazardous waste
About This Item
metody
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
parametry testowe
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: waste water spiked with pesticides
SPE tube/cartridge: Supelclean ENVI-8, glass hardware, 500 mg/6 mL (57107 (500 mg/6 mL); 57108 (1 g/6 mL))
condition: 3 mL methanol; h 2 mL 5% methanol in water
sample addition: load sample at 5 mL/min
drying: 2-3 min with nitrogen stream
elution: 2 x 4 mL hexane:acetone (90:10), drop wise. allow to soak into bed
column: 5% Phenyl Polysiloxane, 15 m × 0.53 mm I.D., 0.50 μm (Supelco equivalent, Equity-5, 28252-U)
oven: 150 °C (0.5 min) to 275 °C (5 min)
inj. temp.: 200 °C
detector: ECD, 300 °C
carrier gas: helium
injection: 1 μL split/splitless (45 sec delay)
przydatność
application for SPE
Zastosowanie
environmental
food and beverages
Zastosowanie
Add 5 mL methanol to 100 mL aqueous hazardous waste. Adjust to pH 5-7 if necessary.
Informacje prawne
Analit
probówka lub płytka SPE
Choose from one of the most recent versions:
Certyfikaty analizy (CoA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Obsługa Klienta
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej