CGEP72TE3
Millipore Express® SHF, Cartridge
cartridge nominal length 20 in. (50 cm), filter diam. 2.7 in. (6.9 cm), Code 7
About This Item
Polecane produkty
Materiały
EPDM rubber seal
polypropylene support
Poziom jakości
zgodność regionalna
meets FDA 21 CFR 177-182 (all component materials)
sterylność
non-sterile
Kompatybilność ze sterylizacją
autoclavable compatible
steam-in-place compatible
linia produktu
EMPROVE® Filter
Millipore Express®
Właściwości
hydrophilic
producent / nazwa handlowa
Millipore Express®
Parametry
≤60 mL/min air diffusion at 2.75 bar (40 psig) (in water)
0.07 bar max. inlet pressure (1 psi) at 135 °C (Reverse)
0.35 bar max. inlet pressure (5 psi) at 135 °C (Forward)
1.7 bar max. inlet pressure (25 psi) at 80 °C (Forward)
1700 mbar max. differential pressure (25 psi) at 80 °C (Forward)
2.07 bar max. inlet pressure (30 psi) at 25 °C (Reverse)
340 mbar max. differential pressure (5 psi) at 135 °C (Forward)
6900 mbar max. differential pressure (100 psi) at 25 °C (Forward)
7 bar max. inlet pressure (100 psi) at 25 °C (Forward)
metody
sterile filtration: suitable
Długość
20 in.
Szerokość
2.7 in.
długość nominalna kartridża
20 in. (50 cm)
śr. filtra
2.7 in. (6.9 cm)
rozmiar filtra
20 in.
powierzchnia filtracyjna
1.08 m2
zanieczyszczenia
≤0.25 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
macierz
Millipore Express® SHF
wielkość porów
0.2 μm pore size
punkt pęcherzykowania
≥4000 mbar (58 psig), air with water
kod wkładu
Code 7
Szukasz podobnych produktów? Odwiedź Przewodnik dotyczący porównywania produktów
Opis ogólny
Opakowanie
Uwaga dotycząca przygotowania
This product was manufactured with a MILLIPORE EXPRESS® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).
Komentarz do analizy
TOC/Conductivity: After sterilization and a controlled water flush of 20 L, samples exhibited <500 ppb TOC per USP 643 and <1.3 μS/cm per USP 645 at 25 °C.
Bacterial Retention: Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 107 CFU/cm2 using ASTM® F838 methodology.
Inne uwagi
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer wetting instructions of Filters with MILLIPORE EXPRESS® Membrane user guide
- Storage Statement: Please refer Filters with MILLIPORE EXPRESS® Membrane user guide
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations
Informacje prawne
Nie możesz znaleźć właściwego produktu?
Wypróbuj nasz Narzędzie selektora produktów.
Certyfikaty analizy (CoA)
Poszukaj Certyfikaty analizy (CoA), wpisując numer partii/serii produktów. Numery serii i partii można znaleźć na etykiecie produktu po słowach „seria” lub „partia”.
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej